Yamato Scientific Vacuum controller Vacuum controller User Manual
Page 25
23
Editing the Setting Point During the Ramp Control
Display After Operation
Key
Explanation
1
•
The ramp control is running.
2
FUNCTION
MENU
•
Press FUNCTION MENU while the
ramp control is running, and the
RAMP TIME edit mode starts.
「rt.」
appears blinking on the indicator.
•
「EDIT」LED is on, indicating that the
edit mode starts.
3
•
Three seconds later, The previous
ramp time appears blinking,
indicating that you can change
values.
4
▼▲
▼▲
▼▲
▼▲
•
Press
「▲」or「▼」, and change the
ramp time you desired. Refer to the
previous chapter
「EDITING THE
SET POINT BEFORE OPERATION
」
in order to input the time.
5
ENTER
•
Press ENTER, and the
「Ramp
Level
」 edit mode starts. 「rL.」
appears blinking on the indicator.
6
•
Three seconds later, the previous
ramp level appears blinking,
indicating that you can change the
values.
7
▼▲
▼▲
▼▲
▼▲
•
Press
「▲」or「▼」, and change the
ramp level you desired. Refer to the
previous chapter in order to input
ramp level.
8
ENTER
•
Press ENTER, and the changed set
point is restored. The ramp control
continues, aiming at the pressure and
the time you have set.
•
「EDIT」LED is off.