Computer – MTS Criterion 60 - Manual User Manual
Page 19

Description
MTS Criterion™ Series 60 Product Manual
Introduction
19
Mixing different brands of hydraulic fluid can affect the system
performance.
Contaminated hydraulic fluid can cause premature wear of the hydraulic
components in your system.
Do not mix different brands of hydraulic fluid. MTS System Corporation
recommends using Exxon Mobile DTE-25.
Computer
The computer is also an integral part of the system. It runs the testing software
which provides full machine control, data acquisition and management, and
advanced data analysis and presentation. MTS has minimized the amount of
custom electronics required for your system, thereby making it flexible and
reliable. This is done by connecting the frame and the computer via standard
USB 2.0 connectors. See the computer configuration below:.
Integrated Operations Platform Properties
P
ARAMETERS
64.305/64.305E
64.605/64.605E/
64.106/64.106E
Pump motor rating
1.5 kW (2.0 hp)
2.2 kW (2.9 hp)
Maximum
continuous pressure
23 MPa (3300 psi)
23 MPa (3300 psi)
Maximum flow
capacity (50 Hz)
3.3 L/min (0.9 gpm)
5.1 L/min (1.3 gpm)
Maximum flow
capacity (60 Hz)
4.0 L/min (1.0 gpm)
6.2 L/min (1.6 gpm)
Filtration
5 microns nominal
5 microns nominal
Reservoir capacity
85 L (22.5 gal)
85 L (22.5 gal)
CAUTION
Monitor and Computer Configuration
D
ESCRIPTION
Monitor
Model
L1710A
Brand
Lenovo Panel
Size
17 inch
Max Resolution
1280×1024
Description
Thinkvision LCD
Industrial PC
Model
ACP-4000