Maintenance and servicing, 1 regular maintenance, 2 cleaning – Memmert ICH C (with CO2 control) Climate chambers User Manual
Page 62

62
D30393 | Date 10/2014
Maintenance and Servicing
8. Maintenance and Servicing
Warning!
Danger due to electric shock. Before doing any maintenance
work, pull out the mains plug.
Warning!
In case of appliances of a certain size, you can get accidentally
locked in, which is life-threatening. Do not climb into the appliance!
Caution!
Danger of cuts due to sharp edges. Always wear gloves when work-
ing in the chamber interior.
8.1 Regular maintenance
Annually:
Annually:
►
Remove dust deposits from the condenser of the cooling unit and the Peltier cooling mod-
ules (see chapters 8.2.4 and 8.2.5).
►
Check the sterile filters and replace them if they are dirty.
►
To guarantee perfect control, we recommend to calibrate the appliance once a year (see
page 52).
Every two years:
Every two years:
►
Replace all sterile filters.
8.2 Cleaning
8.2.1 Interior and metal surfaces
Regular cleaning of the easy-to-clean interior prevents build up of material remains that could
impair the appearance and functionality of the stainless steel chamber over time.
The metal surfaces of the appliance can be cleaned with normal stainless steel cleaning
agents. Make sure that no rusty objects come into contact with the interior or with the stain-
less steel housing. Rust deposits can lead to an infection of the stainless steel. If rust spots
should appear on the surface of the interior due to impurities, the affected area must be im-
mediately cleaned and polished.
8.2.2 Plastic parts
Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic
or solvent-based cleaning agents.
8.2.3 Glass surfaces
Glass surfaces can be cleaned with a commercially available glass cleaner.