Page 60: Menu mode, Display. the display is enlarged and the period

background image

60

D30393 | Date 10/2014

Menu mode

7.8 Protocol

The appliance continually logs all relevant measured values, settings and error messages at

1-minute intervals. The internal log memory is of the continuous memory type. The logging

function cannot be switched off and is always active. The measured data are stored in the

appliance, safe from manipulation. If the power supply is interrupted, the time of the power

failure and voltage recovery are stored in the appliance.
You can export the protocol data for different periods to a USB storage medium via the USB

interface or, via Ethernet, import them to the AtmoCONTROL software for graphical represen-

tation, print-out or storage.

The log memory of the appliance is not modified or deleted by reading it out.

1. Connect the USB storage medium to the

interface on the right side of the control

panel.

2. Activate the protocol. To do so, press

the activation key on the right side of

the

PROTOCOL

display. The display is

enlarged and the period

This Month

au-

tomatically highlighted. To select another

logging period, use the turn control.

UFP 100

Protocol

Last download 12.10.2012 12:00

1 month

SETUP

SOUND

CALIBRATION

LANGUAGE

SETUP

SOUND

CALIBRATION

LANGUAGE

Complete control range

1 week

3. Save your selection by pressing the confir-

mation key. The transfer starts and a

status symbol indicates the progress.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Protocol

letzter Download 12.10.2012 12:00

This Year
Complete control range

This Month

4. As soon as the transfer is complete, a

check mark appears in front of the period

selected. The USB storage medium can

now be removed.

Protocol

letzter Download 25.10.2012 13:20

This Year
Complete control range

This Month

AtmoCONTROL or

read them out via Ethernet, please observe the separate AtmoCONTROL manual.