beautypg.com

Memmert ICH C (with CO2 control) Climate chambers User Manual

Page 43

background image

D30393 | Date 10/2014

43

Malfunctions, warning and error messages

Description

Cause

Action

See

Error message AI E-3 in

the temperature display

TEMP

AI E-3

Set 37.0

°C

37

.4

°C

Temperature monitoring

sensor defective. The oper-

ating sensor takes over the

measurement function.

The appliance can

temporarily be

kept in service

Contact customer

service as soon as

possible

page 2

Error message E-3 in

the temperature display

TEMP

Set 45.0

°C

E-3

°C

Operating and monitoring

sensor defective

Switch off appli-

ance.

Remove the cham-

ber load

Contact customer

service

page 2

Error message E-6 in

the humidity display

HUMIDITY

Set 50

.0%rh

%rh

E-6

Humidity sensor defective

No humidity con-

trol possible

Contact customer

service

page 2

Error message E-5

in the CO

2

display

CO2

Set 15

.0%

%

E-5

CO

2

sensor defective

No CO

2

control

possible

Contact customer

service

page 2

When switching on

the appliance, the start

animation is displayed

in another colour than

white

Cyan

:

Not enough storage

space on the SD card

Red

:

The system files could

not be loaded

Orange

: The

fonts and images could

not be loaded

Contact customer

service

Contact customer

service

Download the firm-

ware update at mem-

mert.com and install it

page 2