Menu mode
7.5.2 Humidity calibration
Humidity control can be adjusted according to customer requirements by means of three
freely selectable balance points. For each selected calibration point, a positive or negative
compensation correction value between –10 % and +10 % can be set (Fig. 28).
For humidity adjustment, you will need a calibrated reference measuring device.
rh 20
+3,0 %
rh 90
-1,5 %
60%
40%
0%
20%
80%
100%
rh 50
+1,5 %
Fact
ory calibr
ation
Fig. 28
Humidity adjustment (example)
Example: Humidity deviation at 60 % should be corrected.
1. Press the activation key to the right of the
CALIB
display. The display is enlarged
and the temperature adjustment option
is automatically highlighted.
UFP 100
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
PROGRAM
TIME
USER ID
PROTOCOL
PROGRAM
TIME
USER ID
PROTOCOL
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
20.0
37.0
5.0
Temperature
Humidity
CO2
2. Turn the turn control until
Humidity
is
Humidity
Humidity
highlighted.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
%rh
%rh
%
0,5
-
%
1,0
+
JUSTIEREN
Temperature
letzte Kalibrierung 12.10.2012 12:00
Cal1
Cal2
10.0
50.0
Humidity
CO2
%rh
%
1,0
+
Cal3
80.0
3. Press the confirmation key repeatedly,
until the calibration point Cal2 is selected.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
%rh
%rh
%
0,5
-
%
1,0
+
JUSTIEREN
Temperature
letzte Kalibrierung 12.10.2012 12:00
Cal1
Cal2
10.0
Humidity
50.0
CO2
%rh
%
1,0
+
Cal3
80.0
4. With the turn control, set the calibration
point Cal2 to 60 % rh.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
%rh
%rh
%
0,5
-
%
1,0
+
JUSTIEREN
Temperature
letzte Kalibrierung 12.10.2012 12:00
Cal1
Cal2
10.0
Humidity
60.0
CO2
%rh
%
1,0
+
Cal3
80.0