Construction and description

2.2 Construction and function

The appliance can heat up the interior to up to 50 °C and cool it down to +10 °C. A compres-

sor is used for cooling.
Humidification of the interior is achieved by evaporating water from a tank at a set rate by

means of a hot-air generator on the rear side of the appliance. The sterile hot air is introduced

into the interior above the fan and mixed with the air current. Humidity is reduced through

condensation, using Peltier cooling modules on the rear side of the appliance. Ice that might

build up due to dehumidification is automatically defrosted in cycles.
Carbone dioxide is introduced into the working chamber via a sterile filter. The turbulence-free

interior ventilation ensures a uniform gas distribution, creating a homogeneous atmosphere.

2.3 Material

For the outer housing, MEMMERT uses stainless steel (Mat.No. 1.4016 – ASTM 430) and for

the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through

its high stability, optimal hygienic properties and corrosion resistance to many (but not all!)

chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with

the materials mentioned. A material resistance table can be requested from the manufacturer.

2.4 Electrical equipment

Operating voltage and current consumption: See nameplate

Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010

Protection type IP 20 acc. to EN 60 529

Interference suppression acc. to EN 55011 class B

Appliance fuse: Safety fuse 250 V/15 A, quick-blow

The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)

2.5 Connections and interfaces

2.5.1 Electrical connection

This appliance is intended for operation on an electrical power system with a system imped-

ance Z

max

of a maximum of 0.292 ohm at the point of transfer (service line). The operator

max

max

must ensure that the appliance is operated only on an electrical power system that meets

these requirements. If necessary, you can ask your local energy supply company what the

system impedance is.
Observe the country-specific regulations when connecting (e.g. in Germany

DIN VDE 0100 with residual current circuit breaker).