Malfunctions, warning and error messages

6.1.3 CO

2

monitoring

Description

Cause

Action

See

Alarm display upper CO

2

alarm limit exceeded

CO2

Set 10

.0%

%

13

.0

Open the door for 30 sec. and

wait to see if the appliance then

steadily adjusts to the setpoint.

If the error occurs again, contact

customer service.

page 2

Alarm display value

below lower CO

2

alarm

limit

CO2

Set 15

.0%

%

12

.0

Check if the door is closed.
Check the correct connection,

valve and level of the gas bottle.

If necessary, connect a new gas

bottle.
If the error occurs again, contact

customer service.

page

22

page 2

6.2 Malfunctions, operating problems and appliance errors

Description

Cause

Action

See

Displays are dark

External power supply

was interrupted

Check the

power supply

page 20

Miniature fuse, appliance

fuse or power module

faulty

Contact customer

service

page 2

Displays cannot be

activated

Appliance locked by USER

ID

Unlock with USER ID

page 61

The appliance is in pro-

gramme, timer or remote

control mode (mode "Write"

or "Write + Alarm")

Wait until the end of

the programme or

timer mode or switch

off the remote control

Displays suddenly look

different

Appliance is in "wrong"

mode

Change to operating

or menu mode by

pressing the

MENU

key

Display T:E-3 in the

temperature display

TEMP

T:E-3

Set 37.0

°C

37

.4

°C

Temperature operating sen-

sor defective. The monitor-

ing sensor takes over the

measurement function.

The appliance can

be temporarily

operated

Contact customer

service as soon as

possible

page 2