Putting into operation

4.3 CO

2

connection

Warning!

Danger of explosion and poisoning when introducing gases/materials

other than CO

2

. Only carbon dioxide may be introduced into the ap-

pliance through the gas connection on the rear of the appliance.

2

2

Warning!

Danger of suffocation. CO

2

can have a suffocating effect in high con-

centrations. In normal operation, the appliance emits small amounts

2

2

of CO

2

to its surroundings. You should therefore ensure that the

room in which it is installed is properly ventilated. Always close the

2

2

stop valve or pressure reducer on the gas bottle if the appliance is

currently not in operation.

Warning!

High concentrations of CO

2

can cause cold burns or frostbite. Avoid

contact with CO

2

gas to the eyes and skin.

2

2

Warning!

Gas bottles may burst or explode at high temperature. Keep the gas

bottles away from open fl ames. Do not store gas bottles at or above

50 °C and ensure that the location is always well-ventilated. Prevent

water from penetrating as well as backfl ow into the gas bottles. It is

essential that you read the safety notes and instructions of the gas

supplier.

Attach the supplied pressure hose to the CO

2

gas bottle (pressure regulator) and to the

“CO2 In” connection on the rear of the

appliance (see Fig. 12). Adjust the pressure

regulator to 0.8 to 1.2 bar.

Notice:

Make sure not to mix up the water and

CO

2

connection.

4.4 Switching on

Switch on the appliance by pressing the On/Off switch

on the front of the appliance ( Fig. 13 ).
The starting process is shown by three animated white

dots

. If the dots have another colour, an error has

. If the dots have another colour, an error has

occurred (see page 43).

After the first start-up, the appliance display is set to

English by default. You can change the language as

described from page 46. However, to get a basic

overview of operating the appliance, you should read

the following chapter first.

Fig. 12

CO

2

connection

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Fig. 13

Switching on the ap-

pliance