beautypg.com

4 date and time, Ufp 100 – Memmert ICH C (with CO2 control) Climate chambers User Manual

Page 50

background image

50

D30393 | Date 10/2014

Menu mode

If humidity/room temperature are particularly high, it is possible that the factory setting for

defrosting, 12 hours, is not sufficient. If this is the case, you should set a more frequent de-

frosting interval, e.g. every 6 hours.
Automatic defrosting is disabled with the setting

Off

. When operating at low temperatures,

this over time causes the cooling unit to ice over. Regular defrosting needs to be carried out in

order to prevent damage to the cooling system.

7.3.7 Remote control

In the setup entry

Remote control

, you can set whether the

appliance should be controlled via remote control and if so,

in which mode. These settings are available:

Off

Read Only

Write

Write+Alarm

When the appliance is in remote control mode, the

symbol appears in the temperature display. In the

symbol appears in the temperature display. In the

settings

Write

and

Write+Alarm

, the appliance cannot be

controlled at the ControlCOCKPIT until the remote control

has been switched off (setting Off) or set to

Read Only

.

In order to use the remote control function, program-

ming skills and special libraries are required.

7.3.8 Gateway

The setup entry

Gateway

is used to connect two networks

Gateway

Gateway

with different protocols.
The gateway is set the same way as the IP address (see page

47).

7.4 Date and Time

In the

Time

display, you can set date and time, time zone and daylight saving time.

Always set the time zone (and summer time yes/no) before you set the date and time.

Avoid changing the set time after that since this can lead to gaps or overlapping when

recording measured values. If you still need to change the time, you should not run a

programme immediately before or after doing so.

1. Activate the time setting. To do so, press

the activation key on the right side of

the

Time

display. The display is enlarged

and the first adjustment option (

Date

)

automatically highlighted.

UFP 100

Date and time

Time
Time zone
Daylight savings

12

.

05

.

2012

12:00
GMT +1

SETUP

SOUND

CALIBRATION

LANGUAGE

SETUP

SOUND

CALIBRATION

LANGUAGE

Date

Setup

Balance

Remote Control

Off

Gateway 192.168.5 .1

+30 %

2/2

23

.2°C

TEMP

Set 180

.0°C

Setup

Balance

Remote Control Off

Gateway

192.168.5 .1

+30 %

2/2