6. Cleaning and Maintenance

When used normally, the Vibratome VT1000 P requires only standard

maintenance.

To ensure trouble-free operation of the instrument over a long period of

time, the following is recommended by Leica:
• Thoroughly clean the instrument on a daily basis.
• From time to time, oil the specimen holder (see

Fig. 4

,

Page 9

), blade

holder and specimen clamp (e.g. after cleaning in the sterilization oven

or using solvents).

• Have the instrument inspected according to the maintenance contract

by a qualified service engineer authorized by Leica.

The intervals depend on how heavily the instrument is used.

All service repairs covered under the Warranty Policy shall be

performed at no expense to the user.

In case of a complete failure of the instrument, follow this procedure:

• First check the power supply on the power socket.
• Then, check the fuse(s) on the rear side of the instrument - see

Chapter 6.3

"Replacing components". For this purpose, it is mandatory

to disconnect the power cable from the power supply.

• If a fuse is defective, you must have identified and corrected the cause

of the burned-out fuse before plugging the power cable back in. If no

fuse is defective, please contact Leica service.

6.2 Maintenance instructions

Only authorized, qualified Leica service personnel may access the

internal components of the instrument for service or repair!