Operation – Leica Biosystems VT1000 P User Manual

Page 38

background image

38

Operating Manual V 1.0 – 01/2010

Be careful when handling the dispos-

able blades or sapphire blade.

The cutting edge is extremely sharp

and can cause serious injuries.

Before inserting a blade, the specimen

to be section should be attached to the

specimen plate (clamped into place or

affixed) to prevent injuries.

Fig. 46

Fig. 47

The blade is held in place by a spring clip (53) in

the blade holder that presses against the bot-

tom of the blade holder.

• First, push the spring clip (53) upwards using

your left thumb (

Fig. 52

).

• To insert the blade, hold it (54) by its dull

ends using your thumb and index finger and

insert it into the blade holder (55) (

Fig. 43

),

then carefully release the spring clip -

Fig. 44

shows the correctly inserted blade.

55

53

54

55

5.

Operation

Inserting the blade

For high-quality sections, a sapphire blade is available (39 053237). This blade fits in

the standard blade holder and provides a superior means for sectioning tissue.

When handling the blades, exercise particular caution to ensure that the blade edge

does not contact any object. Resulting microscopic damage to the blade edge may

cause localized tearing of the specimen. Caution must also be exercised when han-

dling the blades to prevent operator injury from accidental cuts.