Operation – Leica Biosystems VT1000 P User Manual
Page 30
![background image](/manuals/565380/30/background.png)
30
Operating Manual V 1.0 – 01/2010
Fig. 41
Adjustable sectioning window
(continued)
• First, adjust the front limit switch that is on
the front side of the instrument.
• To do so, release the knob (57.
1
) with the
blade fully retracted and move it into the de-
sired position.
• As soon as the switch is in the desired posi-
tion, tighten the knob again so that it cannot
be disengaged during sectioning.
• The rear switch can be moved only while the
instrument is in operation.
• First, turn the knob (57.
2
) counterclockwise
to disengage the switch.
• Then, start the sectioning operation using
the switch for the feed movement of the Vi-
bratome.
• During the feed movement of the blade,
push the switch forward until it reaches the
desired position.
• Finally, tighten the knob to fix the switch in
place.
• You can readjust the sectioning window
whenever new specimen sizes are used.
To adjust the front and rear edge of the section-
ing window, you have to move the internal limit
switch using two rotary knobs (57), which are
on the side of the top plastic cover of the Vi-
bratome. The two rotary knobs can be adjusted
independently of one another.
57.
2
To make this adjustment, the instrument should
be in manual mode.
57.
1
5.
Operation