beautypg.com

Operation – Leica Biosystems VT1000 P User Manual

Page 28

background image

28

Operating Manual V 1.0 – 01/2010

5.

Operation

In some cases, e.g. for large specimens, it is

necessary to move the blade holder (50) on the

mounting bar.
When the width of a large specimen is approxi-

mately the same as the sectioning blade edge

length, the lateral positioning of the sectioning

blade may have to be adjusted to ensure com-

plete sections.

• By loosening the knurled screw (48), you can

move the blade holder sideways along its

axis (49) and thus position it more accurately

relative to the specimen (

Fig. 45

).

48

50

49

Moving the blade holder relative to the specimen

Because the knurled screw (48) is also

used to adjust the blade presentation

angle, the angle may have changed.

Therefore, recheck the blade angle af-

ter this adjustment.

Fig. 39

Adjusting the tilt of the specimen
If the specimen is still not completely horizon-

tal, you can tilt the specimen along another axis

(from side to side).
To do so, tilt the specimen mounting block

slightly when it is clamped in the jaws of the

specimen holder.

Fig. 38

Complete control of the specimen over

all three axes is possible with the di-

rectional specimen receptacle (see

Chap.

7.2

, Optional Accessories).