Leica VT1000 P
4. Setup the Instrument
Fig. 17
4.7 Assembling the specimen clamp
• The specimen clamp (34,
Fig. 16
) is assem-
bled inside the regular black Vibratome tray
(23,
Fig. 15
).
• Insert the cheese head screw (35) into the
bore of the specimen stage (27) and tighten
it using a
size 5
Allen key (37, included in the
scope of delivery).
• When tightening, ensure that the fixture (34)
is parallel to the front edge of the Vibratome
tray (
Fig. 17
).
• Insert the specimen block (36) into the clamp
and clamp it into place using the knurled
screw (39) (
Fig. 18)
.
• The specimen to be sectioned is affixed to
the specimen block using the cyanoacrylate
adhesive provided.
• The purpose of the knurled screw (38) is to
ensure that the
directional specimen fixture
can also be installed (for additional informa-
tion, refer to
Fig. 22
).
Fig. 16
23
27
34
Fig. 15
35
36
37
37
34
Fig. 18
36
38
39
Always remove the blade BEFORE in-
stalling or removing accessories!