5.2.1 Adjusting the blade presentation

angle

• The accessories include a blade angle indi-

cator (46) for this purpose.

• To read the blade angle using this indicator,

hold the angle indicator against the left side

of the blade holder. The notch (47) must be

touching the mounting bar (49) of the blade

holder (50) and the lower edge must be

touching the specimen tray. (

Fig. 40/41

).

• For the adjustment, first unscrew the knurled

screw (48) on the blade holder (

Fig. 41

).

• During the adjustment, it is important that

the bottom edge of the blade angle indicator

is touching the Vibratome tray (23) (arrow in

Fig. 41

). For this purpose, it may be neces-

sary to move the cutting head forward some-

what.

• Now, rotate the blade holder until the small

indicator (51) specifies the desired value on

the scale of the blade angle indicator.

• Retighten the knurled screw so that it cannot

come loose during sectioning.

• For additional information about adjusting

the blade holder, refer to the following sec-

tion.

The presentation angle of the sectioning blade

relative to the plane of the section is adjust-

able to suit operating conditions and specimen

types.

The adjustment can be made by rotating the

blade holder (50) on the mounting bar.