Page 41
41
Leica CM 1850 – Cryostat
9.3
Maintenance
9.3.1 General maintenance
The microtome is virtually maintenance-free. To
ensure a smooth operation of the instrument over
several years we recommend the following:
•
Have the instrument inspected by a qualified
service engineer authorized by us
once a year
.
•
Enter into a service contract at the end of the
warranty period.
For further information, please contact your
local Leica service center.
•
Clean the instrument
every day
.
Once a week:
•
Apply a drop of oil to the plastic coupling
(5,
Fig. 26
)
.
•
Lubricate the specimen cylinder
(1)
:
Push the appropriate coarse feed button to
move the specimen cylinder out to the front
stop position, apply a drop of cryostat oil and
move the specimen cylinder back to the home
position by pressing the appropriate coarse
feed button.
Occasionally, or when required:
•
Lubricate the clamping piece (T-piece)
(2)
on
the microtome base plate and the clamping
lever
(3)
.
•
Lubricate the slot cover
(4)
.
To do so, turn the handwheel to place the
specimen head to the uppermost position and
apply some drops of cryostat oil on to the slot
cover; after that place the specimen head to
Fig. 23
Fig. 24
1
2
3
4
9.
Cleaning, disinfection, maintenance
5