Cleaning, disinfection, maintenance – Leica Biosystems CM1850 UV User Manual
Page 39
39
Leica CM 1850 – Cryostat
9.1
Cleaning
•
Remove frozen section waste from the cryostat with a cold brush every
day.
•
Remove the section waste tray for emptying.
•
Remove the storage shelves and the brush shelf for cleaning.
•
Remove the sliding window by slightly lifting and pulling it to the front
when closed
(see 9.3.6 'Replacement of the lamp')
.
Do not use organic solvents or any other aggressive substances for
cleaning and disinfecting! Only use the cleaning agents specified
in this instruction manual such as Leica Cryofect (alcohol or com-
mon disinfectants based on alcohol)!
•
Drain the cleaning liquid through the hose after the prescribed reagent
time is over and collect it in the waste container
(1)
.
Condensation water produced during defrosting collects in the
waste container. Therefore, check the liquid level regularly and
empty the container if necessary.
9.
Cleaning, disinfection, maintenance
•
Dispose of the waste liquid according to the
waste disposal regulations.
•
To remove the waste container
(1)
, shut off the
tap
(2)
and unscrew the lid
(3)
.
1
2
3
Fig. 22