Technical data – Leica Biosystems CM1850 UV User Manual
Page 12
12
Instructions for use V 2.6 – 10/2012
Type
CM 1850
CM 1850
CM 1850-08
CM 1850-01
Mark of conformity
CUL
-
-
VDE
Nominal voltage (±10%)
120 V AC
220 V AC
230 V AC
240 V AC
Nominal frequency
60 Hz
60 Hz
50 Hz
50 Hz
Power draw
1600 VA
1600 VA
1600 VA
1600 VA
Max. start-up current for 5 sec
45A eff.
45A eff.
45A eff.
45A eff.
Protective class
I
I
I
I
Automatic mains fuse
T12A T1
T12A T1
T10A T1
T10A T1
Pollution degree
①
2
2
2
2
Overvoltage installation category
II
II
II
II
Heat emission (max.)
1600 J/s
1600 J/s
1600 J/s
1600 J/s
Refrigeration system
50 Hz
60 Hz
Cryochamber
Temperature setting range
0 °C to -35 °C (+ 2 K / - 0 K)
0 °C to -35 °C (+ 2 K / - 0 K)
Defrosting
automatic hot gas defrosting,
automatic hot gas defrosting,
temperature controlled
temperature controlled
1 automatic defrost cycle/24 hours,
1 automatic defrost cycle/24 hours,
temperature controlled
temperature controlled
manual defrost cycle
manual defrost cycle
Refrigeration capacity
②
690 W
690 W
Safety factor
3
3
Refrigerant
300 g ±5g refrigerant R 404A
*
300 g ±5g refrigerant R 404A
*
Compressor oil
0.6 l EMKARATE RL-22S, ICI
*
0.6 l EMKARATE RL-22S, ICI
*
Quick freeze shelf
Max. temperature
- 43 °C (+ 3 K/- 2 K)
- 43 °C (+ 3 K/- 2 K)
Number of quick freeze stations 10
10
Defrosting
manual hot gas defrosting
manual hot gas defrosting
time controlled
time controlled
Peltier element
Max. temperature
-60 °C (+5 K)
-60 °C (+5 K)
Number of freezing stations
2
2
*
Defrosting
together with quick freeze shelf
together with quick freeze shelf
Operating temperature range (ambient temperature): 18 °C to 35 °C.
All specifications related to temperature are valid only up to an ambient temperature of 22 °C
and an air humidity lower than 60 %!
3.
Technical data
①
according to IEC-1010, UL 3101
②
according to CECOMAF Liquid temperature 45 °C, evaporation temperature -25 °C