Daily use of the instrument – Leica Biosystems CM1850 UV User Manual
Page 26

26
Instructions for use V 2.6 – 10/2012
7.
Daily use of the instrument
7.1
Specimen freezing
•
Select the appropriate sectioning temperature (cryochamber tempera-
ture) for the sample material (please refer to
7.4 Temperature selection
chart‘).
7.1.1 Quick freeze shelf
The cryochamber has a quick freeze shelf (
5,
Fig. 17) for up to 10 specimen discs.
The temperature of the quick freeze shelf is always lower than the cryochamber
temperature.
•
Cut the specimen to size.
•
Activate the Peltier element
(4)
, if available - it
may take up to 40 seconds until the maximum
refrigeration output is available.
•
Apply enough cryocompound to a specimen
disc at room temperature.
•
Place the specimen on the disc and orient.
•
Place the specimen disc in one of the holes of
the quick freeze shelf and freeze the specimen
at a low temperature.
•
Once the specimen is frozen, insert the
specimen disc in the specimen head
(Fig. 18)
and start sectioning.
7.1.2 Stationary heat extractor
•
Fix the support
(1)
of the heat extractor
(2)
by tightening the 2 screws to
the threaded holes on the left sidewall of the cryochamber and insert the
heat extractor.
•
Lower the heat extractor cylinder on the specimen surface. After ap-
proximately 30 seconds contact, the specimen will be entirely frozen.
•
Once the specimen is frozen, place the heat extractor on the parking
station
(3)
.
1
2
3
4
5
Fig. 17