JUMO 707020 dTRANS T02 - Four-Wire Transmitter Data Sheet User Manual
Page 4

2014-08-08/00379095
Data Sheet 707020
JUMO GmbH & Co. KG • 36035 Fulda, Germany
Page 4/12
Electrical data
Technical data type 707021/..., type 707022/... and type 707025/...
Input for thermocouple
Input for resistance thermometer
Supply voltage
24V DC +10%/-15%
Power consumption
1W
Supply voltage error
± 0.01% per V deviation from 24V
Test voltage
to DIN 61010, Part 1
510V/50Hz, 1min
Isolation
- between input and output
- between input and mains supply
- between output and mains supply
- between input and setup plug
50V
50V
50V
no isolation between input and setup plug
Designation
Range limits
Range
Accuracy
a
Fe-Con L
DIN
43710
-200 to
+900°C
-200 to
+900°C
0.1%
above
-150°C
Fe-Con J
EN
60584
-210 to +1200°C
-200 to +1200°C
0.1%
above
-100°C
Cu-Con U
DIN
43710
-200 to
+600°C
-200 to
+600°C
0.1%
above
-100°C
Cu-Con T
EN
60584
-270 to
+400°C
-200 to
+400°C
0.1%
above
-100°C
NiCr-Ni K
EN
60584
-270 to +1372°C
-200 to +1372°C
0.1%
above
-60°C
NiCr-Con E
EN
60584
-270 to +1000°C
-200 to +1000°C
0.1%
above
-60°C
NiCrSi-NiSi N EN
60584
-270 to +1300°C
-100 to +1300°C
0.1%
above
-80°C
Pt10Rh-Pt
S
EN
60584
-50 to +1768°C
-50 to +1768°C
0.15% above 0°C
Pt13Rh-Pt
R
EN
60584
-50 to +1768°C
-50 to +1768°C
0.15% above 0°C
Pt30Rh-Pt6Rh B
EN
60584
0 —
1820°C
400 —
1820°C
0.15% above 400°C
W3Re-W25Re D
0 —
2495°C
500 —
2495°C
0.15% above 500°C
W5Re-W26Re C
0 —
2320°C
500 —
2320°C
0.15% above 500°C
Shortest span
Type L, J, U, T, K, E, N: 100°C; type S, R, B, D, C: 500°C
Range start/end
freely programmable within the limits in 0.1°C steps
Cold junction
Pt100 internal or external cold junction (adjustable from 0 to 100°C)
Cold junction accuracy
± 1°C
Sampling rate
100msec
Special features
also programmable in °F; input isolated from output
a
The accuracy refers to the maximum range span.
For small ranges, as well as for short spans, the linearization accuracy is reduced.
Designation
Connection circuit
Range limits
Range
Accuracy
Pt 100
EN 60751
2/3-wire
2/3-wire
4-wire
4-wire
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
Pt 100
JIS
2/3-wire
2/3-wire
4-wire
4-wire
-200 to
+649°C
-100 to
+200°C
-200 to
+649°C
-100 to
+200°C
-200 to
+649°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
Pt 500
DIN
2/3-wire
2/3-wire
4-wire
4-wire
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
Pt 1000
DIN
2/3-wire
2/3-wire
4-wire
4-wire
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
-100 to
+200°C
-200 to
+850°C
±0.4°C
±0.8°C
±0.4°C
±0.5°C
Ni 100
2/3-wire
4-wire
-60 to
+180°C
-60 to
+180°C
-60 to
+180°C
±0.8°C
±0.5°C