List of commands – JULABO HT60-M3 High Temperature Circulator User Manual
Page 65
HT60-M2 / HT60-M3
65
11.3. List of commands
When the RS485 interface is used, the instrument address stands in front of each command (Axxx_).
in-commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
version
none
Number of software version (V X.xx)
status
none
Status message, error message (see page 68)
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none
Temperature value registered by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
1 .... 8
Liquid level indication.
in_pv_05
none
Temperature value registered by the safety sensor 1
in_pv_06
none
Temperature value registered by the safety sensor 2
(only HT60)
in_pv_07
none
Temperature value registered by the safety sensor 3
(only HT60)
in_sp_00
none
Working temperature "T"
in_sp_03
none
High temperature warning limit "
".
in_sp_04
none
Low temperature warning limit "
".
in_sp_05
none
Setpoint temperature of the external programmer
(socket 31 - REG+E-PROG) .
in_hil_00
none
Max. cooling power (%).
in_hil_01
none
Max. heating power (%).
in_mode_01
none
Type of setpoint setting
0 = via keypad
2 = via the analog interface E-PROG
in_mode_02
none
Identification type:
0 = no identification
1 = single identification
2 = continual identification