13

2.4. Technical specifications

HT60-M2

HT60-M2 & C.U.

Working temperature range

°C

70 ... 400

40 … 400

Temperature stability external

°C

±0,01 ... ±0.1

Display accuracy

%

±0.5 ±1Digit

Temperature selection

digital

via keypad

indication on LCD DIALOG-DISPLAY (°C/°F)

remote control via PC

indication on monitor

Temperature indication

Multi-Display (LED)

DIALOG-Display (LCD)

Resolution

°C

0.01

Absolute Temperature Calibration

(ATC 1)

°C

±9.99

(ATC 2)

°C

±9.99

Temperature control

ICC - Intelligent Cascade Control,

self-optimizing

Cascade, parameter can be called-in and modified

Heater wattage (at 3x400 V)

kW

7.0

C.U.-cooling unit:

Temperature

Cooling capacity

at 2 l/min flow of cooling water

°C

kW

-----

350 250

150

75

-----

12

8

4

1

Pressure pump:

Max. pressure

max. at 0 liters

bar

0.8 … 1.2

Flow rate

max. at 0 bar

l/min

14 … 18

Electrical connections:

See page 15

Filling volume

litre

2

Overall dimensions (WxDxH)

M2 / M3

cm

25x25x18

25x25x18

Overall dimensions (WxDxH)

HT60

cm

23x23x58

43x23x58

Weight

kg

29

37

Ambient temperature

°C

5 ... 40

Mains power connection

400V/3PNPE/50Hz

V / Hz

365-440V/3PNPE/50Hz & 207-253V/50-60Hz

Current draw

A

10.5 (at 400 V)

10,5 (400V) & 1 (230V)