Remote control, Setup for remote control – JULABO HT60-M3 High Temperature Circulator User Manual
Page 63
HT60-M2 / HT60-M3
63
EXT
Socket for external Pt100 sensor (19)
Pin assignment:
Pin
Signal
1
I+
2
U+
3
U-
4
I-
The shield of the connecting cable is electrically connected to the plug
housing and the sensor tube.
11. Remote control
11.1. Setup for remote control
SERIAL
Select the „Configuration“ submenu and select the option „Setpoint“ to
define the interface (see page 42).
The interface parameters are set by selecting the submenu „Interface“
on the circulator Normally, this is a one-time-only adjustment. (Selecting
and setting menu items, see page 57.)
Factory settings:
RS232
BAUDRATE
4800 bauds
PARITY
even parity
HANDSHAKE
Protocol RTS/CTS
(hardware handshake)
Data bits
7
Stop bit 1
Like all parameters which can be entered through the keypad, interface parameters are stored
in memory even after the circulator is turned off.
4
1
3
2
1
4
2
3
Shield
Plug
Look on
soldering side.
Pt100
9
6
5
1