JULABO HT60-M3 High Temperature Circulator User Manual
Page 15

HT60-M2 / HT60-M3
15
Electrical connections: 
Computer interface 
RS232 or RS485
or Profi
Programmer input
0 - 10 V / 0 - 20 mA
Temperatur recorder outputs
Channel 1 / 2
0 - 10 V /
Channel 3
0 - 20 mA / 4 - 20 mA
Stand-by input
External alarm device
Switching capacity max.
30 W / 40 VA
Switching voltage max.
125 V
∼/−
Switching current max.
1 A
External measurement and controlsensor
Pt100, 4-lead technique
 
 
Safety installations according to IEC 61010-2-010: 
Excess temperature protection 
adjustable from 0 °C to 420 °C
Low liquid level protection
float switch
Liquid level indication
optical 8-graded
Rèpartition par classe selon DIN 12876-1
Classe III FL
Supplementary safety installations
High temperature warning function
optical + audible (in intervals)
Low temperature warning function
optical + audible (in intervals)
Supervision of the working sensor
plausibility control
Reciprocal sensor monitoring between 
working and safety sensors 
difference >100 K
Alarm indication
optical + audible (permanent)
 
Environmental conditions according to IEC 61 010-1: 
• Use only indoor. 
• Altitude up to 2000 m - normal zero. 
• Ambient temperature: see Technical specifications 
• Air humidity: 
Max. rel. humidity 80 % for temperatures up to +31 °C, 
linear decrease down to 50 % relative humidity at a temperature of +40 °C 
• Max. mains fluctuations of ±10 % are permissible. 
• Overvoltage category II 
• Pollution degree 2 
Caution:
The unit is not for use in explosive environment
Protection class according to IEC 60 529: IP31 
The unit corresponds to Class I 
 
Standards for interference resistance according to EN 61326-1 
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class A (industrial 
and commercial range). 
 