Preparations, Setup – JULABO HT60-M3 High Temperature Circulator User Manual
Page 25

HT60-M2 / HT60-M3
25
5. Preparations
5.1. Setup
HT60
CU
-
ESC
P
M2
0
60
180
300
420
EXT
°C
High temperature circulator HT60 as well as
High temperature circulator HT60 with CU cooling
unit:
as a desktop unit
• Place the circulator on an even surface on a pad
made of
non-flammable material.
as a stand-mounted unit
• Place the unit in an vertical position.
• A wall distance of at least 15 cm must be
maintained for ventilation, allowing internal heat to
be conducted away from the unit.
JULABO Order-No.
9790100
C.U. cooling unit
M2 or M3 remote control electronics:
• The unit should be set up at a dry location.
• Place the unit in an upright position and do not
obstruct the ventilation Ambient temperatures
above 40 °C result in a failure of the unit.
• Do not set up the unit in the immediate vicinity of
heat sources and do not expose to sun light.