Menu functions
54
Examples:
lowest temperature value:
10 °C
highest temperature value
210 °C
Fig. shows 200
°C scaled to paper width
rise: 50 mV/°C
lowest temperature value:
197 °C
highest temperature value:
202 °C
Fig. shows 5
°C scaled to paper width
rise: 2000 mV/°C
>EPROG: Current
L Value 0.0°C
H Value 300.0°C
ExtSet 50.0°C
EPROG - Input
This input is necessary when the nominal value is to be determined and
set by an external programmer.
Connect the external programmer to socket (31)
REG+E-PROG of the programmable controller.
• The programmer input of the programmable controller can be
matched to the output signal of the external programmer.
Voltage
voltage input
Current
current input
Select the desired input value with the P-key
and set with enter
.
•
„L Value“ - Setting the LOW value::
Adjust and set the lowest desired working temperature on the
programmer (e.g. 0
°C).
Enter this same temperature on the programmable controller by
pressing the appropriate buttons on the keypad and press enter
to set.
•
„H Value“ - Setting the HIGH value:
Adjust and set the highest desired working temperature on the
programmer (e.g. 300
°C).
Enter this same temperature on the programmable controller by
pressing the appropriate buttons on the keypad and press enter
to set.
• Return to the standard display by pressing escape
.
0 V
0 mA
20 mA
4 mA
10 °C
210 °C
197 °C
202 °C
∆ Τ = 5 K
∆ Τ = 200 K