JULABO HT60-M3 High Temperature Circulator User Manual
Page 33

HT60-M2 / HT60-M3
33
Setp.: 250.00°C
IntAct: 250.00°C
Power 50 %
Control: intern
9. The DIALOG-DISPLAY (LCD) shows the standard indication, the
circulator regulates to the last adjusted setpoint (in the example 250
°C) with the last adjusted control parameters.
The circulator is ready for use!
Stopping the degasifying:
• By operating the Start/Stop-key
the degasifying mode can be
stopped at any time.
• The degasifying mode can also be left via the menu level
.
Proceeding as point 1 – 5.
Overtmp:305.00°C
ExtAct: xxxxxx°C
Power: x %
Control: xxxxxx
Important !
To avoid switching-off due to the excess temperature protection or –
warning both settings should be checked and,.if necessary, adjusted.
• Operate the setpoint-key
.
The MULTI-DISPLAY (LED) shows the present setpoint temperature
(Example: 270.00°C).
(„Safety temperature“ see page 38)
• Operate the setpoint key
.
The DIALOG-DISPLAY (LCD) shows the present setpoint
temperature (Examples: 305.00°C)
(„Warning functions“ see page 37)