JULABO HT60-M3 High Temperature Circulator User Manual
Page 55
HT60-M2 / HT60-M3
55
>EPROG: Current
L Value 0.0°C
H Value 300.0°C
ExtSet 50.0°C
EPROG : 50.00°C
IntAct: 50.00°C
Power: 50 %
Control: intern
3x
Example:
• Setting a temperature of 50 °C on the external programmer!
The value adjusted and set on the external programmer is displayed in
line 4 of the DIALOG-DISPLAY (LCD) for control purposes (Example:
ExtSet: 50.0
°C).
After returning the LCD display to standard display by pressing escape
(„Setpoint“ - see page 42) this value is displayed in line 1
(Example: EPROG 50.00
°C).
EPROG : 152.00°C
IntAct: 50.00°C
Power: 100 %
Control: intern
>EPROG: Current
L Value 20.0°C
H Value 200.0°C
ExtSet 152.0°C
3x
This EPROG input enables the use of different voltage and current
values as program parameters.
•
„L Value“ - Setting the LOW value:
1) Adjust and set the lowest desired value on the voltage or current
source resp. (Example A: 1 V).
2) Assign a lower temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the programmable controller (Example A: 20
°C ) and set
by pressing enter
.
•
„H Value“ - Setting the HIGH value:
1) Adjust and set the highest desired value on the voltage or current
source resp. (Example A: 10 V).
2) Assign an upper temperature threshold value to this adjusted
voltage/current value by pressing the appropriate buttons on the
keypad of the programmable controller
(Example A: 200
°C) and set by pressing enter
.
• Return to the standard display by pressing escape
.
Example B in the diagram serves to illustrate that the end point values
are freely selectable.
Example out of diagram A:
• Adjusting the voltage source for an output of 7.6 V!
Line 4 of the DIALOG-DISPLAY (LCD) shows the externally set setpoint
value. The programmable controller calculates this value from the rise
angle of the two predecided end points (in example A: 7.6 V correspond
to an external setpoint temperature of 152.0
°C ).
After returning the LCD display to standard display by pressing escape
, this value is displayed in line 1
(Example: EPROG 152.00
°C).
Notice:
If this adjustment is not correctly performed at two different points, the setpoint setting will be
incorrect.
20
40
60
80
100
200
300
250
150
°C
1
2 3 4 5 6 7 8 9 10
V
2 4 6 8 10
14
18 20 mA
A
B