Cooling water connection – JULABO HT60-M3 High Temperature Circulator User Manual
Page 16
Operator responsibility – Safety instructions
16
2.5. Cooling water connection
Cooling water pressure (IN / OUT )
max.
4.5 bar
Difference pressure (IN - OUT )
2.0 to 4.5 bar
Rate of flow
typical
2 l/min
Cooling water temperature
<20 °C
Notice:
Danger of corrosion of heat exchanger due to unsuitable quality of cooling water.
•
Due to its high content of lime, hard water is not suitable for cooling and causes scale in
the heat exchanger.
•
Ferrous water or water containing ferrous particles will cause formation of rust even in
heat exchangers made of stainless steel.
•
Chlorinated water will cause pitting corrosion in heat exchangers made of stainless steel.
•
Due to their corrosive characteristics, distilled water and deionized water are unsuitable
and will cause corrosion of the bath.
•
Due to its corrosive characteristics, sea water is not suitable.
•
Due to its microbiological (bacterial) components, which settle in the heat exchanger,
untreated and unpurified river water and water from cooling towers is unsuitable.
•
Avoid particulate matter in cooling water.
•
Avoid putrid water.
Recommended quality of cooling water:
pH
7.5 to 9.0
Sulfate [SO4 2- ]
< 100 ppm
Hydrocarbonate [HCO 3-]/sulfate [SO4 2-]
> 1 ppm
Hardness [Ca 2+, Mg 2+]/[HCO 3-]
> 0.5
°
dH
Alkalinity
60 ppm < [HCO 3-] < 300 ppm
Conductivity
< 500 μS/cm
Chloride (Cl -)
< 50 ppm
Phosphate (PO4 3-)
< 2 ppm
Ammonia (NH3)
< 0.5 ppm
Free chlorine
< 0.5 ppm
Trivalent iron ions (Fe 3+)
< 0.5 ppm
Manganese ions (Mn 2+)
< 0.05 ppm
Carbon dioxide (CO2)
< 10 ppm
Hydrogen sulfide (H2S)
< 50 ppm
Content of oxygen
< 0.1 ppm
Algae growth
impermissible
Suspended solids
impermissible