Control parameters – JULABO HT60-M3 High Temperature Circulator User Manual
Page 44

Menu functions
44
8.2. Control parameters
 
 
 
 
 Configuration
>Control param.
 Profile Start
 Int.Programmer
 Inputs/Outputs
 Limits
 Interface
 Temp.Sensor
Control param.
>Xp = 2.5 K
 Tn = 220 s
 Tv = 1 S
 Xpu= 5.0K
 CoSpeed= 0.0
 DynInt=aperiod
Xp =
When performing an identification for the controlled system (temperature 
applications system) (see page 41), the control parameters Xp, Tn, and 
Tv will be automatically determined and stored. 
Each parameter may be manually set via the keypad if necessary, to 
allow optimum control performance. 
• Press enter
to select the submenu „control parameters“.
• Use the up/down cursor keys
to select the desired
option. A flashing segment indicates that a new value needs to be 
entered. 
• Use the numeral keypad to set the value and then set with enter
(example: Xp = 2.5 °C).
• or at >DynInt<
Press the P-key
to select the parameter and press enter
.
• Press escape
to return to the previous menu level.
Proportional range >Xp< 
The proportional range is the range below the selected temperature 
value in wich the the control circuit reduces the heating power from 100 
% to 0 %. 
Resetting time >Tn< (Integral component) 
Compensation of the remaining control deviation due to proportional 
regulation. An insufficient resetting time may cause instabilities to occur. 
Excessive resetting time will unnecessarily prolong compensation of the 
control difference. 
Lead time >Tv< (Differential component) 
The differential component reduces the control settling time. An 
insufficient lead time will prolong the time required to compensate for 
disturbance effects and cause high overshooting during run-up. An 
excessive lead time could cause instabilities (oscillations) to occur. 
 
 