Preparations, Setup – JULABO HT60-M3 High Temperature Circulator User Manual
Page 25

HT60-M2 / HT60-M3
25
5. Preparations
5.1. Setup
HT60
CU
 
 
 
-
ESC
P
M2
0
60
180
300
420
EXT
°C
High temperature circulator HT60 as well as
High temperature circulator HT60 with CU cooling
unit: 
 
as a desktop unit 
• Place the circulator on an even surface on a pad 
made of
non-flammable material.
as a stand-mounted unit 
• Place the unit in an vertical position. 
• A wall distance of at least 15 cm must be
maintained for ventilation, allowing internal heat to 
be conducted away from the unit. 
 
JULABO Order-No. 
9790100
C.U. cooling unit
 
 
 
 
 
 
 
M2 or M3 remote control electronics: 
• The unit should be set up at a dry location. 
• Place the unit in an upright position and do not 
obstruct the ventilation Ambient temperatures 
above 40 °C result in a failure of the unit. 
• Do not set up the unit in the immediate vicinity of
heat sources and do not expose to sun light.