beautypg.com

JULABO LC 4 Temperature controller User Manual

Page 32

background image

Remote control

32

9.2. List of commands

Befehl Parameter

Description

version none

Number

of

software version (V X.xx)

status

none

Status message, error message (see page 34)

out_mode_05

0

Stop the controller = r OFF

out_mode_05

1

Start the controller

out_mode_02

1

Parameter set 1 effective for control

out_mode_02

2

Parameter set 2 effective for control

out_mode_02

3

Parameter set 3 effective for control

out_mode_02

4

Parameter set 4 effective for control

out_mode_02

5

Parameter set 5 effective for control

in_mode_02

none

Ask for effective parameter set

out_sp_00

xxx.x

Set working temperature

in_sp_00 none

Ask

for

working temperature

in_pv_00

none

Ask for effective value registered by working sensor

in_pv_01

none

Ask for effective heater capacity

out_par_00 xxx.x Set

Xp

of parameter set 1

in_par_00

none

Ask for Xp of parameter set 1

out_par_01

xxx.x

Set Tn of parameter set 1

in_par_01

none

Ask for Tn of parameter set 1

out_par_02

xxxx.x

Set Tv of parameter set 1

in_par_02

none

Ask for Tv of parameter set 1

out_par_03 xxx.x Set

Xp

of parameter set 2

in_par_03

none

Ask for Xp of parameter set 2