beautypg.com

JULABO LC 4 Temperature controller User Manual

Page 27

background image

LC4

27

6.6. Interface parameters: r - br - P - H

RS232C

Whenever the parameter setting of the controller is not conform to
those of the connected personal computer, a modification is
necessary.

Factory setting of the RS232 interface:
Baudrate:

4800

Bauds

Parity:

2 (even)

Handshake: 1 (Hardware handshake)
Data bits:

7

Stop bits:

1

 Press the cursor key

until the desired parameter appears on

the MULTI-DISPLAY (LED).

 Use the edit keys

to set the desired parameter and

then press enter

.


Adjustable interface parameters

REMOTE

0 = keypad control mode

1 = remote control mode via RS232C

BAUDRATE 12 = 1200 bauds

24 = 2400 bauds

48 = 4800 bauds

96 = 9600 bauds

PARITY

0 = no parity

1

=

odd

2

=

even

HANDSHAKE

0 = Protocol Xon/Xoff

(software handshake)

1 = Protocol RTS/CTS

(hardware handshake)

Like all parametes which can be entered through the keypad,
interface parameters are stored in memory even after the controller
is turned off.