Menu functions

26

6.4. Active countercooling: Pc


For applications near the ambient temperature, countercooling
might become necessary.
For this purpose, connect the JULABO MVS solenoid valve
controller to the

/ alarm socket (application example see page

16). The supply of a cooling pulse "Pc" must be activated on the
controller.

Activating the supply of a cooling pulse:

 With the edit keys

change the menu function from

“Pc 0” to "Pc 1" and then press enter

.


Factory setting: "Pc 0"

(no countercooling).

 Whenever countercooling is not necessary, reset the “Pc”

parameter to 0.


6.5. Controlling the heater capacity: h

This control allows the maximum effective heater capacity of the
connected heating device to be adjusted in 10 % steps between 10
% and 100 %. The gradient of the temperature rise function may
thus be controlled to permit gentle warming of thermally sensitive
substances and to prevent over-shooting.



Controlling the heater capacity:

 Using the edit keys

set the desired maximum heater

capacity and then press enter

.

(example: 70 %).


Factory setting: 100 %