JULABO LC 4 Temperature controller User Manual
Page 15

LC4
15
5.3. Connecting a heating device
Caution:
Set up the heating device according to the instructions or securely fix the unit in
the bath tank using appropriate means. Danger of burning and fire!
Connect the power plug to the grounded mains socket (18) on
the rear of the controller.
Caution:
Max. resistive load 1000 W at 115 V / 2000 W at 230 V.
Max. current 5 A at 115 V / 10 A at 230 V
.
5.4. Connecting the temperature sensors
Connect both sensors prior to turning the unit on since an
alarm shutoff will be effected if the sensors are not connected
(see page 28).
Connect the working sensor to the socket “A” (14) and
the safety sensor to socket “S” (15) .
Sensor calibration:
When the controller is first placed into operation or whenever
a sensor is replaced, a working sensor calibration must be
carried out (ATC - see page 23).
Caution:
Place both sensors into the bath medium and securely fix the sensors.
- AWC100 Compact Recirculating Cooler (15 pages)
- CF30 Cryo-Compact Circulators (35 pages)
- CF31 Cryo-Compact Circulators (76 pages)
- ED Heating Immersion Circulator (23 pages)
- ED-5A/B Bath Circulators (24 pages)
- ED-5A Open Bath Circulators (24 pages)
- EH Heating Immersion Circulator (27 pages)
- EH-5 Open Bath Circulator (28 pages)
- F250 Recirculating Cooler (35 pages)
- F12-ED Refrigerated / Heating Circulator (26 pages)
- F12-EH Refrigerated and Heating Circulators (32 pages)
- F12-MA Refrigerated and Heating Circulators (65 pages)
- F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath (69 pages)
- F38-EH Refrigerated and Heating Circulators (30 pages)
- F25-ME Refrigerated and Heating Circulators (66 pages)
- F25-HE Refrigerated and Heating Circulators (86 pages)
- F34-ED Refrigerated / Heating Circulator (25 pages)
- F25-HL Refrigerated Circulators (94 pages)
- FL7006 Recirculating Coolers (34 pages)
- FL2503 Recirculating Coolers (36 pages)
- FL20006 Recirculating Coolers (32 pages)
- FL1201 Recirculating Coolers (33 pages)
- FL300 Recirculating Coolers (48 pages)
- FK30-SL Calibration Baths (88 pages)
- FT200 Immersion Coolers (20 pages)
- FC600_600S Recirculating coolers (37 pages)
- FC1200S Recirculating cooler (45 pages)
- F81-HL Ultra-Low Refrigerated Circulator (93 pages)
- F70-ME Ultra-Low Refrigerated Circulators (64 pages)
- FP52-SL Ultra-Low Refrigerated Circulators (103 pages)
- FP51-SL Ultra-Low Refrigerated Circulators (90 pages)
- FP45-HE Refrigerated and Heating Circulators (79 pages)
- HL-4 Heating Circulators (89 pages)
- HE-4 Heating Circulators (78 pages)
- F81-HL Ultra-Low Refrigerated Circulators (93 pages)
- FT402 Immersion Coolers (24 pages)
- ME-4 Heating Circulators (62 pages)
- ME-31A Visco-Baths (62 pages)
- ME Heating Immersion Circulator (61 pages)
- MB-17 Open Heating Bath Circulators (52 pages)
- MB Heating Immersion Circulator (48 pages)
- Magnum 91 Temperature System (77 pages)
- MA-4 Heating Circulators (51 pages)
- MA Heating Immersion Circulator (50 pages)
- LH 47 Highly Dynamic Temperature Control Systems (76 pages)