JULABO LC 4 Temperature controller User Manual
Page 11

LC4
11
listed in this manual before installing or operating your unit.
Never operate the unit without bath fluid in the bath.
Observe the flash point of the bath medium used. The excess temperature
protection should be set at least 25 °C below the fire point.
Set up the heating device according to the instructions prior to connection to
the controller and ensure secure attachment to the bath.
Danger of burning and fire!
Immerse both temperature sensors in the bath medium and ensure secure
attachment.
Check the filling level of the bath fluid from time to time. The heater must
always be fully covered with the bath fluid!
Never operate damaged or leaking equipment.
Always turn off the unit and disconnect the mains cable from the power
source before performing any service or maintenance procedures, or before
moving the unit.
Sudden jolts or drops may cause damage in the interior of the unit.
Always empty the bath before moving the unit.
Transport the unit with care.
Never operate equipment with damaged mains power cables.
Observe all warning labels.
Never remove warning labels.
Always turn off the unit and disconnect the mains cable from the power
source before cleaning the unit.
Repairs are to be carried out only by qualified service personnel.
Some parts of the bath cover may become extremely warm during
continuous operation. Therefore, exercise particular caution when touching
these parts. Use safety glasses!