Operating procedures – JULABO LC 4 Temperature controller User Manual
Page 14

Operating procedures
14
5. Operating
procedures
5.1. Installation
The unit should be set up at a dry location.
Place
the unit in an upright position and do not obstruct the
ventilation.
A wall distance of at least 10 cm must be maintained for
ventilation, allowing internal heat
to be conducted away from the unit.
If one or more temperature controllers are set up in a
cabinet for example, take care of good ventilation (waste
heat per unit = approx. 60 Watts).
The ambient temperature must not exceed 35 °C. Ambient
temperatures above 35 °C result in a failure of the unit.
Do not set up the unit in the immediate vicinity of heat
sources and do not expose to sun light.
5.2. Power
connection
Caution:
Only connect the unit to a power socket with earthing contact (PE –
protective earth)!
The power supply plug serves as safe disconnecting device from the line and
must be always easily accessible.
Never operate equipment with damaged mains power cables.
Regularly check the mains power cables for material defects (e.g. for cracks).
We disclaim all liability for damage caused by incorrect line voltages!
Make sure that the line voltage and frequency match the
supply voltage specified on the type plate.
Deviations of ±10 % are permissible.