For each parameter set, the control parameters Xp, Tn and Tv
may be individually modified and stored (see page 19).

Preference

:

Quick access to parameters determined during previous
applications.


Factory-preset parameter sets for the following
controlled processes:

Set 1:

Xp = 4 °C
Tn = 150 s
Tv = 10 s

Directly heated liquid bath
Filling volume: 4.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 2000 W

Set 2

Xp = 10 °C
Tn = 300 s
Tv = 25 s

Directly heated liquid bath
Filling volume: 4 litres of Thermal H
Setpoint temperature: 150 °C
Heater capacity: 2000 W

Set 3

Xp = 20 °C

Tn = 1800 s
Tv = 165 s

Heating hood model JH2000 with 2000 ml round bottom
flask
Filling volume: 2 litres of water
Setpoint temperature: 70 °C
Heater capacity: 480 W (stage III)

Set 4

Xp = 35 °C
Tn = 1800 s
Tv = 130 s

Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 150 W (stage II)

Set 5

Xp = 15 °C
Tn = 1550 s
Tv = 120 s

Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of Thermal H
Setpoint temperature: 100 °C