21

5.8.1. Optimization instructions for the PID control parameters

The heat-up curve reveals inappropriate control settings.

t

°C




optimum setting

Inappropriate settings may produce the following heat-
up curves:

t

°C




Xp too low

t

°C




Tv/Tn too low

t

°C



Xp or Tv too high

t

°C



Tv/Tn or Xp too high