Menu functions – JULABO LC 4 Temperature controller User Manual
Page 23

LC4
23
6. Menu
functions
Parameters and values are set via the menu level that may be
selected at any time.
Enter or exit the menu level by pressing the left arrow
and
enter
at the same time.
Using the cursor keys
select the menu functions one
by one.
6.1. Sensor calibration - ATC function
Controller (TT)
Calibration bath (TM)
The ATC (Absolute Temperature Calibration) function serves to
calibrate the working sensor.
Place the working sensor "A" and the temperature sensor of a
thermometer in the calibration bath.
Determine the difference temperature (∆T=TM - TT)
and store as correcting factor (example ∆T = -0.26 °C) as
follows:
With the edit keys
change the menu function from
"Atc 0" to "Atc 1" and then press enter
.
Using the cursor keys
and the edit keys
set the correcting factor (example -0.26 °C) and then press
enter
.
After exiting the menu level, the corrected value (example: 36.7
°C) is indicated on the MULTI-DISPLAY (LED).
Recommendation
:
In case a calibrated temperature measuring instrument is used, the
ATC function allows the controller to be used as testing instrument
according to ISO 9000.