beautypg.com

JULABO LC 4 Temperature controller User Manual

Page 24

background image

Menu functions

24

6.2. Temperature range: HL / LL

The laboratory controller LC4 features the possibility to limit the
working temperature range to protect thermally sensitive
substances.


Setting the upper limit

 With the edit keys

change the menu function from

“HL 0” to "HL 1" and then press enter

.

 Using the cursor keys

and the edit keys

set

the upper limit (example: 250.0 °C) and then press enter

.

HL (High Limit) factory-setting: 350.0 °C

Setting the lower limit

 With the edit keys

change the menu function from “LL

0” to "LL 1" and then press enter

.

 Using the cursor keys

and the edit keys

set

the lower limit (example: -40.0 °C) and then press enter

.

LL (Low Limit) factory-setting: -50.0 °C

6.3. Parameter sets for control: PA


The LC4 provides 5 different parameter sets for controlled
processes.



Selecting a parameter set for control:

 Press the edit keys

to select the desired parameter

set (1 to 5) and then press enter

.

(example: parameter set 3).