beautypg.com

Cleaning / repairing the unit – JULABO FP45-HE Refrigerated and Heating Circulators User Manual

Page 79

background image

HE

79

15. Cleaning / repairing the unit

Caution:

Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.

Prevent humidity from entering into the circulator.

Electrical connections and any other work must be performed by qualified
personnel only.

To maintain the full cooling performance, clean the condenser from time to
time.
• Switch off the unit, disconnect mains power cable.
• Hold the venting grid, pull out and remove.
• Clean the ribbed condenser with a vacuum cleaner.
• Replace the venting grid.
• Switch on the unit.

Cleaning:

For cleaning the bath tank and the immersed parts of the circulator, use
low surface tension water (e.g., soap suds).
Clean the outside of the unit using a wet cloth and low surface tension
water.

The circulator is designed for continuous operation under normal
conditions. Periodic maintenance is not required.

The tank should be filled only with a bath fluid recommended by JULABO.
To avoid contamination, it is essential to change the bath fluid from time to
time.


Repairs

Before asking for a service technician or returning a JULABO
instrument for repair, please contact an authorized JULABO service
station.

When returning the unit:

• Clean the unit in order to avoid any harm to the service personnel.

• Attach a short fault description.

• During transport the unit has to stand upright. Mark the packing

correspondingly.

• When returning a unit, take care of careful and adequate packing.

• JULABO is not responsible for damages that might occur from

insufficient packing.

JULABO reserves the right to carry out technical modifications with repairs
for providing improved performance of a unit.