JULABO FP45-HE Refrigerated and Heating Circulators User Manual
Page 74

Remote control
74
Command
Parameter
Response of circulator
out_par_12
xxx
Tv control parameter of the cascade controller.
0 ... 999
out_par_13
xxx
Maximum internal temperature of the cascade controller.
out_par_14
xxx
Minimum internal temperature of the cascade controller.
out_par_15
xxx
Band limit (upper)
0 ... 200 °C
out_par_16
xxx
Band limit (lower)
0 ... 200 °C
out_hil_00
-xxx
Set the desired maximum cooling power (0 % to 100 %).
Note: Enter the value with a preceding negative sign!
This adjustment only makes sense if FP refrigerating units are
used.
out_hil_01
xxx
Set the desired maximum heating power (10 % to 100 %).
in commands: Asking for parameters or temperature values to be displayed.
Command
Parameter
Response of circulator
version
none
Number of software version (V X.xx)
status
none
Status message, error message (see page 76)
in_pv_00
none
Actual bath temperature.
in_pv_01
none
Heating power being used (%).
in_pv_02
none
Temperature value registered by the external Pt100 sensor.
in_pv_03
none
Temperature value registered by the safety sensor.
in_pv_04
none
Setpoint temperature of the excess temperature protection
in_sp_00
none
Working temperature „SETP 1“
in_sp_01
none
Working temperature „SETP 2“
in_sp_02
none
Working temperature „SETP 3“
in_sp_03
none
High temperature warning limit
„OverTemp“
in_sp_04
none
Low temperature warning limit
„SubTemp“
in_sp_05
none
Setpoint temperature of the external programmer
(REG+E-PROG) .
in_sp_07
none
1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotation speed
after start.
in_sp_08
none
Value of a flowrate sensor connected to the E-Prog input
in_par_00
none
Temperature difference between working sensor and safety
sensor
in_par_01
none
Te - Time constant of the external bath.
in_par_02
none
Si - Internal slope
in_par_03
none
Ti - Time constant of the internal bath.