List of commands – JULABO FP45-HE Refrigerated and Heating Circulators User Manual
Page 73
![background image](/manuals/563230/73/background.png)
HE
73
12.3. List of commands
out commands: Setting temperature values or parameters.
Command
Parameter
Response of circulator
out_mode_01
0
Use working temperature > SETPNT 1<
out_mode_01
1
Use working temperature > SETPNT 2<
out_mode_01
2
Use working temperature > SETPNT 3<
out_mode_02
0
Selftuning „off“.
Temperature control by using the stored parameters.
out_mode_02
1
Selftuning „once“
Single selftuning of controlled system after the next start.
out_mode_02
2
Selftuning „always“
Continual selftuning of controlled system whenever a new
setpoint is to be reached.
out_mode_03
0
Set external programmer input to voltage.
Voltage
0 V ... 10 V
out_mode_03
1
Set external programmer input to current.
Current
0 mA ... 20 mA
out_mode_04
0
Temperature control of internal bath.
out_mode_04
1
External control with Pt100 sensor.
out_mode_05
0
Stop the unit = R –OFF-.
out_mode_05
1
Start the unit.
out_mode_08
0
Set the control dynamics - aperiodic
out_mode_08
1
Set the control dynamics - standard
out_sp_00
xxx.xx
Set working temperature. „SETPNT 1“
out_sp_01
xxx.xx
Set working temperature. „SETPNT 2“
out_sp_02
xxx.xx
Set working temperature. „SETPNT 3“
out_sp_03
xxx.xx
Set high temperature warning limit „OverTemp“
out_sp_04
xxx.xx
Set low temperature warning limit
„SubTemp“
out_sp_06
xxx.xx
Set manipulated variable for the heater via serial interface
-99.99 ... +100 [%]
out_sp_07
x
Set the pump pressure stage. (1 ... 4)
out_par_04
x.x
CoSpeed for external control 0 ... 5.0
out_par_06
xxx
Xp control parameter of the internal controller.
0.1 ... 99.9
out_par_07
xxx
Tn control parameter of the internal controller.
3 ... 9999
out_par_08
xxx
Tv control parameter of the internal controller.
0 ... 999
out_par_09
xxx
Xp control parameter of the cascade controller.
0.1 ... 99.9
out_par_10
xxx
Proportional portion of the cascade controller.
1 ... 99.9
out_par_11
xxx
Tn control parameter of the cascade controller.
3 ... 9999