9.2.
MENU PROGRAM – creation, administration
Menu level 1
1 program
10 sections
The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A
program is composed of individual sections (STEP). The sections are defined
by duration (TSLICE) and target temperature. The target temperature is the
setpoint (SETPNT), which is achieved at the end of a section. The programmer
calculates the temperature ramp from the difference in time and temperature.
STEP
(Nr.)
1
2
4
6
8
SETPNT (°C)
100
180
180
75
75
TIME
(hh.mm)
00:20
00.10
00:20
00:50
00:20
Graph1
25
50
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
75
100
150
200
°C
t
Step 1
2
4
6
8
Sections without set value and time are skipped. They can be defined
retroactively and the integrated into the program.
- AWC100 Compact Recirculating Cooler (15 pages)
- CF30 Cryo-Compact Circulators (35 pages)
- CF31 Cryo-Compact Circulators (76 pages)
- ED Heating Immersion Circulator (23 pages)
- ED-5A/B Bath Circulators (24 pages)
- ED-5A Open Bath Circulators (24 pages)
- EH Heating Immersion Circulator (27 pages)
- EH-5 Open Bath Circulator (28 pages)
- F250 Recirculating Cooler (35 pages)
- F12-ED Refrigerated / Heating Circulator (26 pages)
- F12-EH Refrigerated and Heating Circulators (32 pages)
- F12-MA Refrigerated and Heating Circulators (65 pages)
- F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath (69 pages)
- F38-EH Refrigerated and Heating Circulators (30 pages)
- F25-ME Refrigerated and Heating Circulators (66 pages)
- F25-HE Refrigerated and Heating Circulators (86 pages)
- F34-ED Refrigerated / Heating Circulator (25 pages)
- F25-HL Refrigerated Circulators (94 pages)
- FL7006 Recirculating Coolers (34 pages)
- FL2503 Recirculating Coolers (36 pages)
- FL20006 Recirculating Coolers (32 pages)
- FL1201 Recirculating Coolers (33 pages)
- FL300 Recirculating Coolers (48 pages)
- FK30-SL Calibration Baths (88 pages)
- FT200 Immersion Coolers (20 pages)
- FC600_600S Recirculating coolers (37 pages)
- FC1200S Recirculating cooler (45 pages)
- F81-HL Ultra-Low Refrigerated Circulator (93 pages)
- F70-ME Ultra-Low Refrigerated Circulators (64 pages)
- FP52-SL Ultra-Low Refrigerated Circulators (103 pages)
- FP51-SL Ultra-Low Refrigerated Circulators (90 pages)
- HL-4 Heating Circulators (89 pages)
- HE-4 Heating Circulators (78 pages)
- F81-HL Ultra-Low Refrigerated Circulators (93 pages)
- FT402 Immersion Coolers (24 pages)
- ME-4 Heating Circulators (62 pages)
- ME-31A Visco-Baths (62 pages)
- ME Heating Immersion Circulator (61 pages)
- MB-17 Open Heating Bath Circulators (52 pages)
- MB Heating Immersion Circulator (48 pages)
- Magnum 91 Temperature System (77 pages)
- MA-4 Heating Circulators (51 pages)
- MA Heating Immersion Circulator (50 pages)
- LH 47 Highly Dynamic Temperature Control Systems (76 pages)