Technical specifications – JULABO FP45-HE Refrigerated and Heating Circulators User Manual
Page 11

HE
11
2.4.
Technical specifications
FP45-HE
Working temperature range
°C
-42 ... 200
Temperature stability
°C
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -30
0.85 0.7 0.42 0.08
Refrigerant
R-404A
Overall dimensions
(WxDxH)
cm
38x58x69
Bath opening
(WxL)
cm
23x26
Bath depth
cm
20
Filling volume
liters
18 ... 26
Weight
kg
53
Mains power connection
230 V/50 Hz
V/Hz
207-253/50-60
Current draw
(at 230 V)
A
13
Mains power connection
230 V/60 Hz
V/ Hz
207-253/50-60
Current draw
(at 230 V)
A
13
All measurements have been carried out at:
rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.
HE
Temperature selection
digital
via keypad
indication on VFD COMFORT-DISPLAY
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY
Resolution
°C
0.01
ATC3
INT/EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage (at 230 V)
kW
2.0
Heater wattage (at 115 V)
kW
1.0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
Max. pressure
at 0 liters
bar
0.7
Max. suction
at 0 liters
bar
0.4
Electrical connections:
See page 12
Ambient temperature
°C
5 … 40