2.4.

Technical specifications

FP45-HE

Working temperature range

°C

-42 ... 200

Temperature stability

°C

±0.01

Cooling capacity

Medium: ethanol

°C
kW

+20 0 -20 -30
0.85 0.7 0.42 0.08

Refrigerant

R-404A

Overall dimensions

(WxDxH)

cm

38x58x69

Bath opening

(WxL)

cm

23x26

Bath depth

cm

20

Filling volume

liters

18 ... 26

Weight

kg

53

Mains power connection

230 V/50 Hz

V/Hz

207-253/50-60

Current draw

(at 230 V)

A

13

Mains power connection

230 V/60 Hz

V/ Hz

207-253/50-60

Current draw

(at 230 V)

A

13



All measurements have been carried out at:
rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.

HE

Temperature selection

digital

via keypad

indication on VFD COMFORT-DISPLAY

remote control via PC

indication on monitor

Temperature indication

VFD COMFORT-DISPLAY

Resolution

°C

0.01

ATC3

INT/EXT

°C

±3 / ±9

Temperature control

ICC - Intelligent Cascade Control

Heater wattage (at 230 V)

kW

2.0

Heater wattage (at 115 V)

kW

1.0

Electronically adj. pump capacity

stages

1 ... 4

Flow rate

at 0 bar

l/min

22 ... 26

Max. pressure

at 0 liters

bar

0.7

Max. suction

at 0 liters

bar

0.4

Electrical connections:

See page 12

Ambient temperature

°C

5 … 40