6.2.2.

Switching on the Cooling Machine

Switching on:

• Switch on the cooling machine using the switch (1b) .








Control of the cooling machine:

With the mains switch (1b) turned on, the circulator automatically
switches the cooling machine off and on.
• It is switched off, if:

- the actual working temperature is increased by >30 °C (cooling is
not required).
- the heater operates at full power (>800 W) for longer than 5
minutes.

• It is switched on, if:

- cooling is necessary for maintaining the bath temperature.
After switch-off, the cooling machine automatically switches on only
after a delay of 5 minutes for protecting the cooling compressor.

To save energy, turn off the cooling machine with the mains switch