Electrical connections – JULABO FP45-HE Refrigerated and Heating Circulators User Manual
Page 70
Electrical connections
70
11. Electrical connections
Notice:
Use shielded cables only.
The shield of the connecting cable is electrically connected to the plug housing.
The unit ensures safe operation if connecting cables with a maximum length of 3 m
are used. The use of longer cables does not affect proper performance of the unit,
however external interferences may have a negative impact on safe operation (e.g.
cellular phones).
4
1
3
2
ext. Pt100
1
4
2
3
Shield
Plug
Look on
soldering side.
Pt100
Socket for external Pt100 sensor
Pin assignment:
Pin
Signal
1
I+
2
U+
3
U-
4
I-
The shield of the connecting cable is electrically connected to the plug
housing and the sensor tube.
RS232 serial interface
This port can be used to connect a computer with an RS232 cable for
remote control of the circulator.
9
6
5
1
RS232C
Pin assignments RS232:
Pin 2
RxD
Receive Data
Pin 3
TxD
Transmit Data
Pin 5
0 V
Signal GND
Pin 7
RTS
Request to send
Pin 8
CTS
Clear to send
Pin 1; 4; 6, 9 Reserved - do not use!
RS232 interface cable
Circulator (9-pol)
PC (9-pol)
Pin 2 RxD
⇔
Pin 3 TxD
Pin 3 TxD
⇔
Pin 2 RxD
Pin 5 GND
⇔
Pin 5 GND
Pin 7 RTS
⇔
Pin 8 CTS
Pin 8 CTS
⇔
Pin 7 RTS
Accessories:
Order No.
Description
8 980 073
RS232 interface cable 9-pol./9-pol. , 2,5 m
8 900 110
USB interface adapter cable