9.4.6.
Setting of clock and date
The internal real time clock allows starting a program any time. The clock 
is set to the local mean time (MEZ) at the factory. 
If the unit is operated in a different time zone, the clock can be 
adjusted in this menu. 
Change summer/winter time in this menu
 
 
9.4.7.
RESET – Factory settings
A Reset will return all values to factory setting except for date and time.
A RESET can be effected in the >OFF< mode only.
Switch off the circulator by pressing the key
and call up the
menu CONFIGURATION.
 
 
 
 
9.5.
MENU CONTROL – Control characteristics and parameters
Menu level 1
The circulator is qualified for internal and external temperature control 
The switchover is carried out in the menu >C-TYPE< .(INT or EXT). 
For external temperature control and measurement connect a Pt100 
external sensor to the socket at the rear of the circulator. 
Press the
key if a parameter is to be retained. Correction function
for parameters or values (prior to OK)
Level 2
Parameter level
or
The parameter flashes, switch by pressing
and
The control type can be adjusted in the -OFF-
mode only.
Depending on the adjustment only the active 
parameters are displayed. 
or
or
The parameter flashes, switch by pressing
and
OFF - no selftuning.
ONCE - single selftuning (factory setting)
ALWAYS - continual selftuning.